Definite diagnosis | AGSCS | Total | G-test | Total |
---|
Positive | Negative | Positive | Negative |
---|
Positive | 76 | 6 | 82 | 71 | 11 | 82 |
Negative | 12 | 104 | 116 | 20 | 96 | 116 |
Total | 88 | 110 | 198 | 91 | 107 | 198 |
- The 198 serum sample from human were detected using AGSCS and G-test. The sensitivity of AGSCS and G-test were 92.68% and 86.58%, the specificity of both were 89.65% and 82.75%, respectively. The accuracy of AGSCS and G-test were 90.9% and 84.34% respectively. The AGSCS showed better detectability for patients with IFI than G-test (2 = 3.936, p = 0.0473).